For Display setup, analog outputs and setpoints see the section "General Setup".
CO
One point Calibration
2
A
180.4
hPa
A
26.1
°C
Calibrate Sensor
Channel A CO2
H
180.4
hPa
A
26.1
°C
CO2 Calibration
Type = 1 Point
H
137.5
hPa
A
26.1
°C
Press ENTER when
Sensor is in Buffer 1
154.5
hPa
A
26.1
°C
A Point1 = 7.00 pH
A
CO2 = 7.07 pH
H
154.5
hPa
A
26.1
°C
pH S=100.0%
Z=7.048pH
Save Adjust
Select ADJUST to save the calibration values and the successful Calibration is confirmed
on the display. If you select calibrate, the values will not be taken. If an ISM Sensor is
used, the Calibration data will be stored in the calibration history.
(PATH: Cal)
While in measurement mode press the key [CAL]. Press the
Conductivity calibration and press the [ENTER] key to call the sub function.
Select 1 point and press [ENTER]
Place the sensor in the buffer solution and press the [ENTER] key to start the calibra-
tion.
The display shows the buffer the transmitter has recognized (Point 1) and the
measured value.
As soon as the drift conditions have stabilized the display changes to show the slope
calibration factor S and the offset calibration factor Z.
After a successful calibration, the calibration values are taken over (Adjust) or were
aborted (Calibrate or Abort).