Herunterladen Inhalt Inhalt Diese Seite drucken

M400 Type 1 Cond Ind Only: Conductivity Calibration For Inductive Sensors - Mettler Toledo M400 Kurzanleitung

Vorschau ausblenden Andere Handbücher für M400:
Inhaltsverzeichnis

Werbung

Verfügbare Sprachen
  • DE

Verfügbare Sprachen

  • DEUTSCH, seite 27
10 M400 Type 1 Cond Ind only:
Conductivity Calibration for inductive Sensors
For Display setup, analog outputs and setpoints see the section "General Setup".
Cond Ind Zero point Calibration
1.25
A
µS/cm
25.00
A
°C
Calibrate Sensor
Channel A Conductivity m
1.25
A
µS/cm
25.00
A
°C
Cal Compensation
Standard
m
40.5
H
mS/cm
23.9
A
°C
Conductivity Calibration
Type = Zero Point
1.035
A
mS/cm
21.9
A
°C
A Point 1 = 0.000 mS/cm
A
C = 1.035 mS/cm
While in measurement mode press the key [CAL]. Press the m or . key to select the
CONDUCTIVITY CALIBRATION and press the [ENTER] key to call the sub function.
Select the kind of calibration (standard, linear or n. water).
Select ZERO POINT and press [ENTER]. Remove the sensor and dry it.
u
Press the [ENTER] key again to go on with the calibration.
As soon as the conditions have stabilized (or [ENTER] pressed in the manual mode)
the display changes to show the multiplier and adder.
u
After a successful calibration, the calibration values are taken over and stored in the
cal history (Adjust), only stored in the cal history (Calibrate) or the calibration can
be aborted (Abort).
(PATH: Cal)
19

Werbung

Inhaltsverzeichnis
loading

Inhaltsverzeichnis