VPAP technical specifications
Pressure and
flow state
Mode reading
Physical
Air filter
Sound
DECLARED
DUAL-NUMBER
NOISE EMISSION
VALUES in
accordance with
ISO 4871: 1996
6
Operating pressure range (measured at the mask): 4–20 cm H
2–25 cm H
O (VPAP)
2
Maximum single fault steady state pressure: 30 cm H
exceeded for > 6 sec; 40 cm H
Pressure measurement tolerance: ±0.5 cm H
Flow measurement tolerance: ±6 L/min or 10% of reading, whichever is greater,
at 0 to 150 L/min positive flow
CPAP mode
Set Pressure: 4 to 20 cm H
S, ST and T modes
IPAP: 4 to 25 cm H
O; EPAP: 2 to 25 cm H
2
Nominal dimensions (L x W x H): 153 mm x 140 mm x 86 mm
Weight: 835 g
Housing construction: Flame retardant engineering thermoplastic
Air outlet: 22 mm conical air outlet (complies with ISO 5356-1:2004)
Hypoallergenic air filter: Acrylic and polypropylene fibers in a polypropylene
carrier
Standard air filter: Polyester non-woven fiber
Pressure level (CPAP mode)
With SlimLine tube:
With Standard tube:
With either SlimLine
or Standard tube and
H5i:
O (VPAP) - if pressure exceeded for >1 sec
2
O
2
O
2
24 dBA as measured according to ISO 17510-1:2002
26 dBA with uncertainty of 2 dBA as measured according
to ISO 17510-1:2007
24 dBA as measured according to ISO 17510-1:2002
27 dBA with uncertainty of 2 dBA as measured according
to ISO 17510-1:2007
27 dBA as measured according to ISO 17510-1:2002
28 dBA with uncertainty of 2 dBA as measured according
to ISO 17510-1:2007
O (CPAP),
2
O (CPAP) - if pressure
2
O ± 4% of the measured reading
2