Herunterladen Inhalt Inhalt Diese Seite drucken

ResMed VPAP ST Bedienungsanleitung Seite 5

Positive airway pressure device
Vorschau ausblenden Andere Handbücher für VPAP ST:
Inhaltsverzeichnis

Werbung

Verfügbare Sprachen

Verfügbare Sprachen

Problem/possible cause
The following message is NOT displayed on the LCD after you try to update the settings using
Settings updated successfully, press any key
the SD card:
The settings were not updated.
General technical specifications
90W power supply unit
Power supply
Input range: 100–240V, 50–60Hz, 110V, 400Hz nominal for aircraft use
Typical power consumption: 70W (80VA)
Maximum power consumption: 110W (120VA)
30W power supply unit
Input range: 100–240V, 50–60Hz, 110V, 400Hz nominal for aircraft use
Typical power consumption: 20W (30VA)
Maximum power consumption: 36W (75VA)
90W DC/DC converter
Nominal inputs: 12V, 24V
Typical power consumption: 70W (80VA)
Maximum power consumption: 110W (120VA)
Environmental
Operating temperature: +5°C to +35°C
conditions
Operating humidity: 10 to 95% non-condensing
Operating altitude: Sea level to 2,591m
Storage and transport temperature: -20°C to +60°C
Storage and transport humidity: 10 to 95% non-condensing
Aircraft use
ResMed confirms that the device/s meets the Federal Aviation Administration
(FAA) requirements (RTCA/DO-160F, section 21, category M) for all phases of air
travel.
Electromagnetic
Product complies with all applicable electromagnetic compatibility requirements
(EMC) according to IEC60601-1-2, for residential, commercial and light industry
compatibility
environments. Information regarding the electromagnetic emissions and
immunity of this ResMed device can be found on www.resmed.com, on the
Products page under Service and Support. Click on the PDF file for your
language.
Class II (double insulation), Type BF
IEC 60601-1
classification
Solution
Contact your clinician/service provider immediately.