Problem/possible cause
There is excessive leak.
Note: If Leak Alert is enabled, an
audible alert is activated and a high
leak message is displayed.
SmartStart/Stop is disabled.
Note: SmartStart/Stop is not available
if Leak Alert is enabled.
Device does not stop when you remove your mask
SmartStart/Stop is disabled.
Note: SmartStart/Stop is not available
if Leak Alert is enabled.
SmartStart/Stop is enabled but the device does not stop automatically when you remove your
mask
Incompatible mask system being
used.
The patient is using a nasal pillows
mask or pediatric mask with a set
pressure less than 7 cm H
Pressure rises inappropriately
Talking, coughing or breathing in an
unusual manner.
Mask cushion is buzzing against the
skin.
Cushion seated incorrectly causing
excessive leak.
Displays error message:
Device has been left in a hot
environment.
Solution
Adjust position of mask and headgear.
Air tubing not connected properly. Connect firmly at both
ends.
Enable SmartStart/Stop.
Enable SmartStart/Stop.
Only use equipment recommended by ResMed.
Disable SmartStart/Stop.
O.
2
Avoid talking with a nasal mask on, and breathe as normally
as possible.
Adjust the headgear.
Adjust headgear or re-fit cushion.
High temperature fault, refer to user manual
Allow to cool before re-use. Disconnect the power cord and
then reconnect it to restart the device.