Fit the new batteries making sure they are
get
inserted with the poles in the right direction, as
eter
indicated on the battery compartment.
ing
Now replace the cover checking it is firmly
ped
closed.
ain
Warning: replace with 1.5V AAA alkaline
batteries only (not rechargeable versions).
ars.
Never use other types of batteries to prevent
the
fire and explosion risks.
Warning: the battery can explode if not
handled with care. Do not recharge, pull apart,
overheat or set the battery on fire.
Keep the battery out of reach of children.
ure,
To dispose of the thermometer, please follow
out
the instructions provided by EC Directive
not
2002/96/EC (see page 22); to dispose of the
h in
battery, please follow the instructions provided
by EC Directive 2006/66/EC (see page 22).
ing
hes
TECHNICAL SPECIFICATIONS
1. Conforms with EN 60601-1 and EN 60601-
and
1-2 Standard requirements.
2. Developed in conformance with EN
OT
12470-5 Standard requirements for Clinical
in
Thermometers – Section 5: IR thermometer
om
performances (with maximum device)
0°C
3. Temperature reading range:
"Fever" Mode: 34~42.2°C (93,2~108°F);
"Ambient Mode": 10~40°C (50°F~104°F).
RY
4. Accuracy level:
ine
"Fever" Mode: +/-0.2°C(0.4°F) between
35.5°C (95.6°F) and 42.0°C (107.6°F) and
ery
+/-0.3°C (0.5°F) beyond this range;
arts
"Ambient Mode": +/-1°C(1.8°F)
l it
5. Clinical repeatability: 0.18°C (< 1 year old),
am
0.17°C (1~5 years old), 0.15°C (>5 years old)
6. Operating range:
(50 °F ~104 °F); relative humidity < 85%
7. Transport conditions: -20~65°C (-4 °F ~104
°F); relative humidity < 90%
4
Classified to EN 60601-1 International
Standards :
1. Internally powered appliance.
2.
3. Continuous operation mode
CE0197: conforms to the EC 2007/47/EC
Medical Equipment Standard and following
amendments.
NOTE:
1. The thermometer was produced in
compliance with all standards currently in
force and meets all magnetic compatibility
protection requirements.
However, it is necessary to take certain
EMC precautions and prepare and use the
thermometer following the EMC instructions
provided in this document.
2. Portable and mobile radio communication
devices can affect the correct functioning
of electro-medical devices. In the presence
of electromagnetic disturbance, it is highly
recommended to repeat the temperature
reading if there is any doubt about the results.
Portable and mobile RF communication
equipment can affect medical electrical
equipment such as this product. This device
complies with CISPR 11 group 1, Class B.
During measurement it is recommended to
keep at least 1.5 meter separation between
this device and the surrounding electrical
devices. If interference occurs please increase
the separation further or turn off the electrical
10 ~ 40°C
equipment in the vicinity.
For detailed description regarding the EMC
precautions you need to take please contact
our authorized representative or us on address
as given in manual.
19
Applied part: BF type.