Use

General

Make all vacuum manifold and electrical connections and refer to
Turbo-V pump instruction manual before to operating the Turbo-V
controller.
To avoid injury to personnel and damage to the equipment, if the pump is
WARNING!
laying on a table make sure it is steady. Never operate the Turbo-V pump if
the pump inlet is not connected to the system or blanked off.
The pump is factory set with Argon process gas. To avoid pump failures, do not
CAUTION!
select Air if the pump is used with heavy gases.
The input signal P1 connector should be left in position including the shipping
NOTE
links if no external connections are made. The forepump and the Turbo-V pump
can be switched on at the same time.
When the pressure at the Foreline of the Turbo-V 2000HT pump is higher than 0.5
NOTE
mbar, the Soft Start must be deselected.
Turbo-V 2000HT Controller User Manual / 87-900-913-01 (O)
Technical Information

Use

265/306