the [Kd-735] is intended for use in an electromagnetic environment in
which radiated rf disturbances are controlled. the customer or the user of
the [Kd-735] can help prevent electromagnetic interference by maintaining
a minimum distance between portable and mobile rf communications
equipment (transmitters) and the [Kd-735] as recommended below,
according to the maximum output power of the communications
equipment.
Rated maximum
output power
of transmitter
W
0,01
0,1
1
10
100
for transmitters rated at a maximum output power not listed above, the recommended separation distance d in
meters (m) can be determined using the equation applicable to the frequency of the transmitter, where P is
the maximum output power rating of the transmitter in watts (W) according to the transmitter manufacturer.
Note 1: at 80 mhz and 800 mhz, the separation distance for the higher frequency range applies.
Note 2: these guidelines may not apply in all situations. electromagnetic propagation is affected
by absorption and reflection from structures, objects and people.
Separation distance according to frequency of transmitter m
150 kHz to 80 MHz
d = 1.2 √ P
0,12
0,38
1,2
3,8
12
ENGLISH
80 MHz to 800 MHz
d = 1.2 √ P
0,12
0,38
1,2
3,8
12
800 MHz to 2,5 GHz
d = 2.3 √ P
0,23
0,73
2,3
7,3
23