33
t
able 2
For all ME EQUIPMENt and ME SYStEMS
Guidance and manufacturer's declaration - electromagnetic immunity
the [Kd-735] is intended for use in the electromagnetic environment
specified below. the customer or the user of the [Kd-735] should assure that
it is used in such an environment.
IMMUNItY test
electrostatic
discharge (esd)
IeC 61000-4-2
power frequency
(50/60 hz)
magnetic field
IeC 61000-4-8
Note: ut is the a.c. mains voltage prior to application of the test level.
able 3
For ME EQUIPMENt and ME SYStEMS that are not LIFE-SUPPORtING
Guidance and manufacturer's declaration - electromagnetic immunity
the [Kd-735] is intended for use in the electromagnetic environment specified
below. the customer or the user of the [Kd-735] should assure that it is used
in such an environment.
IMMUNItY test
IEC 60601
Compliance
test level
level
± 6 kv
contact ±
± 6 kv contact ±
8 kv air
8 kv air
3 a/m
3 a/m
IEC 60601
Compliance
test level
level
ENGLISH
Electromagnetic environment -
guidance
floors should be wood, concrete
or ceramic tile. If floors are
covered with synthetic material,
the relative humidity should be at
least 30%
power frequency magnetic fields
should be at levels characteristic of
a typical location in a typical
commercial or hospital
environment
Electromagnetic environment -
guidance