1. Remove four (4) #10 bolts securing Junction Box top cover. Remove cover and set aside
2. Observing the upstream to downstream orientation, pull Channel-1 and Channel-2 UP
FUT1010 IP65 NEMA 4X & IP66 NEMA 7 Gas Quick Start
Operating Instructions, 1/2013, A5E02639183-AC
bolts.
CH 4-Sensor to Junction Box Gland
CH 3-Sensor to Junction Box Gland
Top Cover #10 bolts (4)
Temperature Sensor Terminal Board
TB1
CH 2-Sensor to Junction Box Gland
Transducer Channel F-Connectors
Junction Box without top cover
CH4 - DN to Transmitter
Figure 3-6
Sensor Cable Installation
and DN Sensor cables into the corresponding CH-1 and CH-2 Junction Box glands.
NOTICE
Incorrect Bend Radius
If bend radius is incorrect damage to cables may result.
When connecting sensor cables inside Junction Box cable bend radius should not
exceed bend tighter than 8 cm (3 in.).
CH 4 - UP to Transmitter
CH 2 - DN to Transmitter
CH 2 - UP to Transmitter
Temperature Sensor Gland
CH 1 - DN to Transmitter
CH 1 - UP to Transmitter
CH 3 - DN to Transmitter
CH 3 UP to Transmitter
CH 1-Sensor to Junction Box Gland
Connecting
3.2 Sensor Wiring