Sagittal plane
Heel height: Effective heel height (shoe heel height - sole thickness of fore­
   
foot) + 5 mm
Exterior foot rotation: approx. 5°
   
   
Anterior placement, middle of the prosthetic foot to the alignment reference
line: 30 mm
   
Connect the prosthetic foot and prosthetic socket using the chosen adapters.
The instructions for use of the adapters must be observed.
   
Determine the centre of the prosthetic socket with the 50:50 gauge. Align the
prosthetic socket centrally to the alignment reference line.
Socket flexion: Individual residual limb flexion + 5°
Frontal plane
Alignment reference line of prosthetic foot: between big toe and second
   
toe
Alignment reference line of prosthetic socket: along the lateral patella
edge
   
Observe the abduction or adduction position.
TF bench alignment
► Observe the information in the prosthetic knee joint instructions for use.
5.1.2 Static Alignment
Ottobock recommends checking the alignment of the prosthesis using
the L.A.S.A.R. Posture and adapting it as needed.
If necessary, the alignment recommendations (TF modular leg pros­
theses: 646F219*, TT modular leg prostheses: 646F336*) may be
requested from Ottobock.
5.1.3 Dynamic Trial Fitting
Adapt the setup of the prosthesis in the frontal plane and the sagittal
plane (e.g. by making angle or slide adjustments) so that correct heel
contact and optimal roll-over are ensured.
TT fittings:Make sure that physiological knee movement is achieved
after the heel makes impact and the leg begins to bear weight.
Remove the adapter cover from the prosthetic foot after completion of
the dynamic trial fitting and gait training exercises.
5.2 Optional: Installing the foam cover
A connecting element (such as a connection plate, connection cap or foam
connection cap) provides a detachable connection between the foam cover
and the prosthetic foot.
Bench alignment process