2 Intended use
2.1 Indications for use
The product is to be used solely for lower limb prosthetic fittings.
2.2 Area of application
The product is recommended for mobility grade 2 (restricted
outdoor walker) and mobility grade 3 (unrestricted outdoor walk
m°
kg
er).
Size [cm]
Max. body weight [kg]
2.3 Environmental conditions
Allowable environmental conditions
Temperature range for use: -10 °C to +60°C
Allowable relative humidity 0 % to 90 %, non-condensing
Unallowable environmental conditions
Mechanical vibrations or impacts
Perspiration, urine, fresh water, salt water, acids
Dust, sand, highly hygroscopic particles (e.g. talcum)
2.4 Service Life
This product has been tested by the manufacturer with 2 million load cycles
according to ISO 22675. Depending on the patient's activity level, this cor
responds to a service life of 2 to 3 years.
3 Safety
3.1 Explanation of warning symbols
Warning regarding possible risks of accident or injury.
CAUTION
Warning regarding possible technical damage.
NOTICE
3.2 General safety instructions
CAUTION
Exceeding the service life and reuse on another patient
Risk of injury due to loss of functionality as well as damage to the product
10
22 to 25
75
26 to 30
100