U93
Parameters
Calibration Date
Calibration Initials
Calibration Period
(days)
Measurement location
ID
1)
Typical values for an HBM U93/2 kN type force transducer
When creating the Bridge Sensor template, the manufacturer defines the
physical measured quantity and the physical unit.
The available unit for the particular measured quantity is specified in the
IEEE Standard. For the measured quantity of force, this is the unit N.
At the time of creating the template it is also necessary to choose between the
options "Full Precision", "mV/V" and "μV/V" for the accuracy of the
characteristic transducer curve mapped in TEDS.
HBM always opts for "Full Precision" here, in order to be able to use full digital
resolution. This choice is also recommended to users who program the TEDS
memory themselves.
For more extensive information about TEDS, look in the TEDS Operating
Manuals on the Internet at www.hbm.com/TEDS
A2081-2.1 en/de/fr
1)
Value
Unit
1−Jun−2006
HBM
730
days
0
Change
re-
quires
rights
to level:
CAL
Date of the last calibration or cre-
ation of the test certificate (if no cal-
ibration carried out), or of the stor-
age of the TEDS data (if only nomi-
nal (rated) values from the data
sheet were used). Format: day−
Abbreviations for the months: Jan,
Feb, Mar, Apr, May, Jun, Jul, Aug,
Sep, Oct, Nov, Dec.
CAL
Initials of the calibrator or calibration
laboratory concerned.
CAL
Time before recalibration, calculated
from the date specified under Cal-
USR
Identification number for the mea-
suring point. Can be assigned ac-
cording to the application. Possible
values: a number from 0 to 2047. If
that is not enough, the HBM Chan-
nel Comment template can also be
used for this purpose.
Explanation
month−year.
ibration Date.