6.2.4 Tolerance value for
calibration stick setting
6.2.5 Linearity control intervall
setting
This menu allows you to enter the tolerance limits for the calibration stick.
Normally, the tolerance value is ± 4, but the calibration stick value tolerances vary from
country to country. If necessary, the tolerance values can be changed in this menu.
This menu allows you to enter the linearity control interval in hours (!).
8640 hours
= 360 days
4320 hours
= 180 days
180 hours
= 7 ½ days
i The settings vary from country to country!
65