Read and set service-
parameters
6.2.1 Write-protection for Date/Time
setting
The service switch S3 must be set (slide at bottom position) in order to change a
parameter.
The switch is located behind the screwed-on cover panel on the front side of the RTM.
i You must trigger a reset after setting the service switch.
S1
Bootstrap
S2
Download
S3
Service
The "Read and set service-parameters" menu contains the following sub-menus:
– Write-protection for Date/Time setting (Chapter 6.2.1)
– Linearity control setting (Chapter 6.2.2)
– Calibration stick value setting (Chapter 6.2.3)
– Tolerance value for calibration stick setting (Chapter 6.2.4)
– Linearity control interval setting (Chapter 6.2.5)
– Service interval setting (Chapter 6.2.6)
– Adjustment calibration interval setting (Chapter 6.2.7)
– Date of last calibration setting (Chapter 6.2.8)
– Zero calibration interval setting (Chapter 6.2.9)
– Automatic switch to idle interval setting (Chapter 6.2.10)
– Self test check valve and ventilators setting (Chapter 6.2.11)
– Limit for optical contamination setting (Chapter 6.2.12)
– Minimum temperature of chamber setting (Chapter 6.2.13)
– Display opacity value setting (Chapter 6.2.14)
– Pressure correction setting (Chapter 6.2.15)
– Temperature correction setting (Chapter 6.2.16)
This menu allows you to activate or deactivate write protection for the date and time.
S1 S2 S3
63