RS 232 interface

3.2
Reference conditions
(calibration conditions)
50
The interface is used to connect to the processor unit of the emission system analyzer
or emission analysis module and for connection to a service PC.
Pin assignments of the 16-pin AMP socket
Pin
Signal
10
TXD
11
RXD
12
Shield
Interface parameters
Baudrate
Parity
Word length
Stop bit
Handshake
Sync. signal
All control functions, measured value requests, measured value transfers, calibration
functions and device or error status signals are transferred or triggered via the serial
interface.
Nominal operating conditions
Ambient temperature
Atmospheric pressure fluctuations referred to
the atmospheric pressure at the operating location
Relative humidity of the ambient air
Range in which accuracy
must be maintained
The data complies with the following requirements:
– German National Standards Laboratory (PTB):
PTB-A 18.9, 2nd Draft dated 06.06.1994
– EC:
ECE Regulation No. 24
– Italy:
CUNA Standard NC 005-04 dated 21.05.1982 and
CUNA Standard NC 005-05 dated 21.05.1982
– Switzerland:
VAMV dated 20.10.1992
– ISO:
ISO 3173 (Draft)
ISO 11614
– France:
French Standard
Classification Code NF R 10 - 025 - 1 (dated 1995)
NF R 10 - 025 -2 (dated 1995)
NF R 10 - 025 -3 (dated 1995)
NF R 10 - 025 -4 (dated 1995)
9600 baud
none
8 bits
1 bit
none
STX / EXT
+2°C to +40°C
± 25 hPa
≤ 90%
no condensation
+5°C to +45°C