Herunterladen Inhalt Inhalt Diese Seite drucken

Hameg HMS 3000 Handbuch Seite 33

3ghz spektrumanalysator
Inhaltsverzeichnis

Werbung

Verfügbare Sprachen

Verfügbare Sprachen

1 GHz Spectrum Analyzer HMS1000, HMS1010 (with TG)
[3 GHz Spectrum Analyzer HMS3000, HMS3010 (with TG)]
All data valid at 23 °C after 30 minute warm-up
Frequency
Frequency range:
HMS1000, HMS1010
100 kHz...1 GHz
HMS3000, HMS3010
100 kHz...3 GHz
Temperature stability:
±2 ppm (0...30 °C)
Aging:
±1 ppm/year
Frequency counter (from SW 2.0):
Resolution
1 Hz
Accuracy
±(Frequency x tolerance of reference)
Span setting range:
HMS1000, HMS1010
0 Hz (zero span) and 1 kHz...1 GHz
HMS3000, HMS3010
0 Hz (zero span) and 100 Hz...3 GHz
Spectral purity, SSB phase noise:
30 kHz from carrier
<-85 dBc/Hz
(500 MHz, +20...30 °C)
100 kHz from carrier
<-100 dBc/Hz
(500M Hz, +20...30 °C)
1 MHz from carrier
<-120 dBc/Hz
(500MHz, +20...30 °C)
Sweep time:
Span = 0 Hz
20 ms...100 s
Span > 0 Hz
20 ms...1000 s, min. 20 ms/600 MHz
Resolution bandwidths (-3 dB):
HMS1000, HMS1010
1 kHz...1 MHz in 1–3 steps, 200 kHz
HMS3000, HMS3010
100 Hz...1 MHz in 1–3 steps, 200 kHz
Tolerance:
±5 % typ.
≤300kHz
1MHz
±10 % typ.
Resolution bandwidths (-6 dB):
HMS1000, HMS1010
9 kHz, 120 kHz, 1 MHz
HMS3000, HMS3010
200 Hz, 9 kHz, 120 kHz, 1 MHz
Video bandwidths:
10 Hz...1 MHz in 1–3 steps
Amplitude
Display range:
Average noise level displayed up to +20 dBm
Amplitude measurement
range:
Typ. -114...+20 dBm
Max. permissible DC
at HF input:
80 V
Max. power at HF input:
20 dBm, 30 dBm for max. 3 Min.
Intermodulation free range:
TOI products, 2x -20 dBm
66 dB typ. (typ. +13 dBm third-order
(-10 dBm ref. level)
intercept)
(at distance between
signals ≤2 MHz)
60 dB typ. (+10 dBm TOI)
(at distance between
signals >2 MHz)
66 dB typ. (typ. +13 dBm TOI)
DANL (Displayed average noise level):
(RBW 1 kHz, VBW 10 Hz,
ref. level ≤-30 dBm
10 MHz...1 GHz resp. 3 GHz) -105 dBm, typ. -114 dBm
With Preamp.
HMS1000/HMS1010: -125 dBm typ. (1 kHz RBW)
HMS3000/HMS3010: -135 dBm typ. (100 Hz RBW)
Inherent spurious:
(ref. level ≤-20 dBm,
f >30 MHz, RBW ≤100 kHz) <-80 dBm
Input related spurious:
(Mixer level ≤-40 dBm,
carrier offset >1 MHz)
-70 dBc typ., -55 dBc (2...3 GHz)
2
nd
harmonic receive frequency
(mixer level -40 dBm):
-60 dBc typ.
Level display:
Reference level
-80...+20 dBm in 1 dB steps
Display range
100 dB, 50 dB, 20 dB, 10 dB, linear
Logarithmic display scaling dBm, dBµV, dBmV
Linear display scaling
Percentage of reference level (from SW 2.0)
Measured curves:
1 curve and 1 memory curve
Trace mathematics:
A-B (curve-stored curve), B-A
Detectors:
Auto-, Min-, Max-Peak, Sample, RMS,
Average, Quasi-Peak
Failure of level display:
<1.5 dB, typ. 0.5 dB
(ref. level to ref. level-50 dB, 20...30 °C)
HMS3000/3010E/121010 · C&E · Subject to change without notice · © HAMEG Instruments GmbH
HAMEG Instruments GmbH · Industriestr. 6 · D-63533 Mainhausen · Tel +49 (0) 6182 8000 · Fax +49 (0) 6182 800100 · www.hameg.com · info@hameg.com
Marker/Deltamarker
Number of marker:
8
Marker functions:
Peak, next peak, minimum, center
= marker, frequency, reference level
= marker level, all marker on peak
Marker displays:
Normal (level, lin. & log.), delta marker, noise
marker, (frequency) counter (from SW 2.0)
Inputs/Outputs
HF Input
N socket
Input Impedance:
50 Ω
VSWR
<1.5 typ.
(10 MHz...1 GHz/3 GHz):
Output tracking generator:
(HMS1010/HMS3010)
N socket
Output Impedance:
50 Ω
Frequency range:
5 MHz...1 GHz [3 GHz]
Output level:
-20...0 dBm, in 1 dB steps
Trigger and external
reference input:
BNC female, selectable
Trigger voltage
TTL
Reference frequency
10 MHz
Essential level (50 Ω)
10 dBm
Supply output for field probes: 6 V
dc
Audio output (Phone):
3.5 mm DIN jack
Demodulation
AM and FM (internal speaker)
Miscellaneous
Display:
16.5 cm (6.5") TFT Color VGA Display
Save/Recall memory:
10 complete device settings
Trigger:
Free run, Video Trigger (from SW 2.0),
Single Trigger, external Trigger
Interfaces:
Dual-Interface USB/RS-232 (HO720),
USB-Stick (frontside), USB-Printer
(rear side), DVI-D for ext. monitor
Power supply:
105...253 V, 50/60 Hz, CAT II
Power consumption:
Max. 40 Watt at 230 V, 50 Hz
Protection class:
Safety class I (EN61010-1)
Operating temperature:
+5...+40 °C
Storage temperature:
-20...+70 °C
Rel. humidity:
5...80 % (non condensing)
Dimensions (W x H x D):
285 x 175 x 220 mm
Weight:
3.6 kg
Accessories supplied: Line cord, Operating manual, CD, HZ21 Adapter plug, N
plug to BNC socket (2x HMS1010/3010)
Recommended accessories:
HO730
Dual-Interface Ethernet/USB
HO740
Interface IEEE-488 (GPIB), galvanically isolated
HO3011
Preamplifier -135 dBm DANL (100 Hz RBW)
HZ13
Interface cable (USB) 1.8 m
HZ14
Interface cable (serial) 1:1
HZ20
Adapter, BNC to 4 mm banana
HZ33
Test cable 50 Ω, BNC/BNC, 0.5 m
HZ34
Test cable 50 Ω, BNC/BNC, 1 m
HZ46
4RU 19'' Rackmount Kit
HZ72
GPIB-Cable 2 m
HZ99
Carrying Case for protection and transport
HZ520
Plug-in Antenna with BNC connection
HZ525
50 Ω-Termination, N plug
HZ530
Near-Field Probe Set 1 GHz for EMV diagnostics
HZ540/550
Near-Field Probe Set 3 GHz for EMV diagnostics
HZ540L/550L Near-Field Probe Set 3 GHz for EMV diagnostics
HZ547
3 GHz VSWR Bridge for HMS1010, HMS3010
HZ560
Transient limiter
HZ575
75/50 Ω Converter
HZO30
active probe 1 GHz (0.9 pF, 1 MΩ, including many accessories)
®
· DQS-certified in accordance with DIN EN ISO 9001:2008, Reg.-No.: 071040 QM08
S p e c i f i c a t i o n s
, max. 100 mA (2.5 mm DIN jack)
Subject to change without notice