165X Users
Change #1
On page 29, replace the second Note with the following:
Model 1651 does not allow the Auto selection.
On page 33, replace step 2, with the following:
2. Press 1 to select the RCD current rating (10, 30,100, 300 or 500 mA).
On page 49, under Loop and Line Impedance, (Z
Maximum Test Current @ 400 V
Maximum Test Current @ 230 V
Replace the third table with the following:
Range
20 Ω
200 Ω
2000 Ω
*Valid for resistance of neutral circuit <20 Ω
**Valid for mains voltage >200 V
11/03
Note
Resolution
0,01 Ω
0,1 Ω
1 Ω
), add the following to the table:
I
20 A for 10 ms
12 A for 10 ms
Accuracy*
± (3% + 10 digits)
± 3%
± 6% **
1