EN
Manufacturer
Model
Type
Mains voltage ( 10%)
Mains frequency
Connected load
Current consumption
Cooling medium
Max. capacity
Allowed density
Speed (RPM)
Force (RCF)
Kinetic energy
Obligatory inspection (BGR 500)
Ambient conditions (EN / IEC 61010-1)
Set-up site
Altitude
Ambient temperature
Humidity
Excess-voltage category
(IEC 60364-4-443)
Pollution degree
Device protection class
EMC
Emitted interference,
Interference immunity
Noise level (dependent on rotor)
Dimensions
Width
Depth
Height
Weight
34/121
Andreas Hettich GmbH & Co. KG
D-78532 Tuttlingen
UNIVERSAL 320 R
1406-20
200-240 V 1
240 V 1
50 Hz
60 Hz
800 VA
4.0 A
Up to 2000 m above sea level
Maximum relative humidity 80% for temperatures up to 31°C, linearly
decreasing to 50% relative humidity at 40°C.
Not suitable for use in explosion-endangered areas.
EN / IEC 61326-1, Class B
approx. 52 kg
1406-21
115-127 V 1
60 Hz
950VA
8.0 A
R 404A
4 x 100 ml
3
1.2 kg/dm
15000
21382
8600 Nm
no
Indoors only
5°C to 35°C
2
FCC Class B
58 dB(A)
401 mm
695 mm
346 mm