The "beurer Health Manager" app must be active to allow
data transfer.
If your smartphone has a protective cover, remove this to ensure
that there is no interference during the transfer. Begin the data
transfer in the "beurer Health Manager" app.

9. Cleaning and storing the device and cuff

• Clean the device and cuff carefully using a slightly damp cloth
only.
• Do not use any cleaning agents or solvents.
• Under no circumstances hold the device and cuff under wa-
ter, as this can cause liquid to enter and damage the device
and cuff.
• If you store the device and cuff, do not place heavy objects on
the device and cuff. Remove the batteries. The cuff line should
not be bent sharply.

10. Error messages/troubleshooting

In the event of errors, the error message
display.
Error messages may appear if:
• it was not possible to record the pulse:
• you move or speak during the measurement:
• the cuff is fastened too tightly or loosely:
• errors occur during the measurement:
• the pump pressure is higher than 300 mmHg:
• the batteries are almost empty
the data could not be sent via Bluetooth
In such cases, repeat the measurement and/or data transfer.
_ appears on the
1;
2;
3;
4;
5;
6;
:
.
®
30
Ensure that the cuff line is correctly attached and that you do not
move or speak during the measurement.

11. Technical specifications

Model No.
BM 54
Measurement me-
Oscillometric, non-invasive blood pres-
thod
sure measurement on the upper arm
Measurement range
Cuff pressure 0 – 250 mmHg,
Systolic 50 – 250 mmHg,
Diastolic 30 – 200 mmHg,
Pulse 40 – 180 beats/minute
Display accuracy
Systolic ± 3 mmHg,
Diastolic ± 3 mmHg,
Pulse ± 5 % of the value shown
Measurement inac-
Max. permissible standard deviation
curacy
according to clinical testing:
Systolic 8 mmHg /
Diastolic 8 mmHg
Memory
2 x 60 memory spaces
Dimensions
L 139 mm x W 94 mm x H 48 mm
Weight
349 g (without batteries, with cuff)
Cuff size
22 to 44 cm
Permissible operating
+ 5°C to + 40°C, 15-93% relative humi-
conditions
dity (non-condensing)
Permissible storage
-25°C to + 70°C, ≤ 93% relative humidity,
conditions
700 – 1060 hPa ambient pressure
Power supply
4x 1.5 V
AAA batteries