15. CLEANING

Alc.
The surface of the floor stand can be easily kept clean by wiping it with
a damp cloth. Any common cleaning agents may be used.
Should disinfection be necessary, the use of an agent that is either
dissolved in water or max. 20% alcohol is recommended.

16. MAINTENANCE

Caution!
All models may only be serviced if the power cord has
been unplugged.
When assembling the KaWe examination light and once per year
thereafter, the mounting pin and the groove are to be lightly greased
with non-acidic grease.

Adjusting the spring tension

If the light no longer stays in the desired position, proceed as follows:
• Remove the light body. Loosen the threaded pin 4 M5x10 in the cap
ring of the telescoping tube until the light body can be pulled up out
of the stand post
• Push the telescoping tube of the stand post all the way down until
it is at its lowest point. Inside of the tube you will see an adjustment
screw.
• Turn the adjustment screw with a screwdriver as follows:
20%
The floor stands do not contain any hazardous materials.
At the end of the product life, the components of the KaWe exami-
nation light are to be disposed of properly. Ensure that the materials
are sorted carefully according to type. The floor stand parts are to be
disposed of appropriately according to their type of material.
a. If the lamp body does not stay in place and keeps sink-
ing: turn the screw clockwise. This increases the tension.
b. If the lamp body is hard to move upward: Turn the screw
counter-clockwise. This decreases the tension.

17. DISPOSAL

31