2.8.2 Rated data

For rated data please observe section 2.7.2

2.8.3 Special conditions

In type of protection intrinsically safe apparatus, the
analytical probes (InPro 6000 Series with InFit 76X/Y
and InTrac 7XX/YY Series housings) equipment is
designated with the following specific conditions of
use.
1. The oxygen sensor shall be installed in compli-
ance with the enclosure, mounting, spacing and
segregation requirements of the ultimate applica-
tion, including a tool removable cover, and is suit-
able for use with FM Approved InFit 76X/Y and In-
Trac 7XX/YY Series Housings.
2. Process temperature no greater than +130 ºC.
3. Maximum permissible working pressure is
12 barg (174 psig).
4. Potential Electrostatic Charging Hazard –
To prevent the risk of electrostatic sparking, the
non-metallic surface should only be cleaned with
a damp cloth.
5. Enclosures containing titanium constitute a poten-
tial risk of ignition by impact or friction. Care must
be taken into account during installation and use
to prevent impact or friction.