Storage and cleaning

Keep the unit in the storage case when not in use. Clean the product with a soft, slightly
damp cloth. Do not wet the cuff or attempt to clean the cuff with water. Do not wash
or dry clean the cuff. Do not store the product in a place exposed to direct sunlight or
high humidity. Remove the batteries if the unit will not be used for 3 months or longer.
Always replace all of the batteries with new ones at the same time.
Specifications
Model:
BPW4100
Operation principle:
Oscillometric method
Display:
Liquid crystal display
Measuring range:
0-280mmHg (cuff pressure)
Blood pressure measurement:
Systolic pressure: 60-280mmHg
Diastolic pressure: 30-250mmHg
Pulse rate:
40-180/min
Laboratory accuracy:
Cuff pressure: ±3mmHg
Pulse rate: ±5%
Inflation:
Automatic, soft adapted to blood pressure
Power supply:
2 batteries, type LR03(AAA)
Electric shock protection:
Internal power supply appliance type BF
Operating environment:
Temperature +10 °C ~ +40 °C
Humidity 15 ~ 85%RH (noncondensing)
Atmospheric pressure 105 kPa ~ 80 kPa
Transport and
Temperature -20 °C ~ +60 °C
storage environment
Humidity 15 ~ 85%RH (noncondensing)
Atmospheric pressure 105 kPa ~ 80 kPa
Cuff:
Suitable for wrist circumference of 13.5 to 21.5cm
Accessories:
Batteries(2*AAA)
If device is not used within specified temperature and humidity ranges the technical
accuracy can not be ensured.
12
What to do if...
Problem
Reason
Battery icon is flashing
• Batteries are flat.
• Memory readings can be
called up, but measure-
ments are not possible.
Err 1
Cuff is not wrapped on
the wrist well (too loose).
Err 2
• Arm moved during
inflation measurement.
• Systolic pressure is
above 280mmHg
No display
No information on the
display screen.
Measurements appear
Blood pressure varies
too high or too low
constantly during the
day. Many factors such
as stress, time of day,
talking, coughing, laugh-
ing, moving etc. when
measuring will influence
the reading.
Prior to measurement,
Positioning system (A) is
the positioning system
deactivated.
(A) on the display does
not work.
Solution
Insert new batteries.
Rewrap the cuff tighter
on the wrist.
Please do not move arm
or speak when you take a
measurement.
Check the battery
installation for proper
installation of the
batteries' polarities
Please read "key rules for
accurate blood pressure
measurements" and
"taking a measurement"
Refer to the chapter
«select mode» to activate
the positioning system
(A).
13