Herunterladen Inhalt Inhalt Diese Seite drucken

Procedures (Mv Or Volts Output); Procedures (Ma Output); Procedures (Idos Umm) - GE Druck DPI 832 Benutzerhandbuch

Inhaltsverzeichnis

Werbung

Verfügbare Sprachen
  • DE

Verfügbare Sprachen

  • DEUTSCH, seite 19
Calibration - Procedures (mV or Volts output)
1. Connect the instrument to the calibration equipment
(Figure 2).
2. Let the equipment get to a stable temperature
(minimum: 5 minutes since the last power on).
3. Use the calibration menu (Table 9) to do a two-point
calibration (Zero and +FS). The display shows the
applicable instructions to complete the calibration.
4. To make sure that the calibration is correct, select the
applicable Volts output task and range (Table 2).
5. Then set the output values that are applicable to the
calibration:
• mV: 0, 30, 60, 90, 120
OR
• Volts (V): 0, 3, 6, 9, 12
6. Make sure that the error is in the specified limits
(Table 13 or Table 14).
Table 13: mV output error limits
Output
Calibrator
mV
error
(mV)
0
0.00005
30
0.000425
60
0.0008
90
0.001175
120
0.00098
Table 14: Volts (V) output error limits
Output
Calibrator
V
error
(V)
0
0.00000005
3
0.0000175
6
0.00003
9
0.00005
12
0.000134
K393 Issue 3
Permitted
DPI 832 error
(mV)
0.01
0.02
0.03
0.03
0.04
Permitted
DPI 832 error
(V)
0.001
0.002
0.002
0.002
0.002
Calibration - Procedures (mA output)
1. Connect the instrument to the calibration equipment
(Figure 3).
2. Let the equipment get to a stable temperature
(minimum: 5 minutes since the last power on).
3. Use the calibration menu (Table 9) to do a two-point
calibration (Zero and +FS). The display shows the
applicable instructions to complete the calibration.
4. To make sure that the calibration is correct, select the
applicable mA output task (Table 2) and set these
output values:
• mA: 0.1, 4, 12, 20, 24
5. Make sure that the error is in the specified limits
(Table 15).
Table 15: mA output error limits
Output
Calibrator
mA
error
(mA)
0.1
0.000006
4
0.00020
12
0.0014
20
0.00185
24
0.0023
Calibration - Procedures (IDOS UMM)
Refer to the user manual for the IDOS UMM.
When the calibration is complete, the instrument
automatically sets a new calibration date in the UMM.
Permitted
DPI 832 error
(mA)
0.001
0.001
0.001
0.002
0.002
11
[EN] English -

Werbung

Inhaltsverzeichnis
loading

Inhaltsverzeichnis