Installation
WA R N I N G
Electrical hazards and fire hazard. Use only the supplied power cable.
Only qualified experts may perform the tasks described in this section of the
manual, in compliance with all locally applicable safety regulations.
N O T I C E
Use only earthed sockets for the connection of this device to the power supply.
If it is not clear whether the sockets are earthed, have this checked by a qualified
electrician.
In addition to supplying power, the power plug also serves to isolate the device
quickly from the mains where necessary.
The entire measurement system has two power plugs (measurement device and
refrigerator). During the disconnection from the mains, it must be made sure that
the correct power plug is pulled (e.g. labeling of the sockets).
This is recommended for long-term disuse and can prevent potential dangers in
the event of a fault.
Therefore make sure that the sockets to which the device is connected are easy
to reach by each user at all times.
N O T I C E
As the refrigerator itself is not provided with a fuse, it may only be connected to a
10 A fused CEE socket.
Please only perform the following work steps as a
pair:
1.
Route the mains connection cable of the refrigerator to the outside.
2.
Set the thermostat knob to level 5–6.
3.
Connect the tray drain hose (8/11).
4.
Place the measurement device on the refrigerator and attach with
the side lock screws.
Figure 3 Refrigerator connection