The special shape of the overflow vessel used has important
functions and some particular advantages:
Sample removal in the device, directly at the sample feed, and thus
a small dead time
Further routing of the excess sample and use in any downstream
devices (for example phosphate or
nitrate photometer)
Air bubble separation
Sample reserve for short-term failure of sample supply
Figure 10 Dosing pump head during roller exchange
1
Sample removal
3
2
Sample feed
4
It is only attached with a retaining clamp and can be easily pulled out for
cleaning purposes.
The incoming and outgoing hoses are connected with hose connectors
and can also be easily pulled out. To clean the overflow vessel and the
hoses, a weak sodium chloride solution (chlorine bleach) is sufficient.
With extreme dirt or strong bacteria growth, the hoses should be
completely exchanged.
Further routing of the excess sample
Retaining clamp
Stirrer vessel
The stirrer vessel provides the fast and thorough mixing of the sample
(null and standard solution) and the reagents A and B. The stirrer vessel
can be dismantled conveniently and without special aids.
For assembly, first insert the stirrer disk and O-ring and then screw the
seal plug hand tight.
The dosing hoses are inserted in the stirrer vessel via hollow screws.
Figure 11 Stirrer vessel
1
Sample (null and
standard solution)
2
Reagent A
3
Reagent B
4
M4 hexagon socket head screw
Observe the following during maintenance and care:
5
Stirrer disk
6
O-ring
7
Sealing plug