Error messages

All notes and error messages show in the bottom left of the display.
Detailed error details can be found in normal cases in the [STATUS]
menu – there, the errors can also be acknowledged.
Error
Definition
SIGNAL LOW
Cuvette not filled
Optics or cuvette
dirty
Lamp failure
Lamp voltage too low
Amplifier defective/not
adjusted
Concentration too high
SIGNAL HIGH
Cuvette not filled
Lamp voltage too high
Amplifier defective/not
adjusted
TEST
Invalid or not plausible
SETTINGS
data in the device data
memory
Frequent occurrence
points to an empty
buffer battery
SET
Time and/or date have
TIME/DATE
invalid settings and
were reset to 1/1/93,
15:00
LEAK IN THE
The moisture sensor
SYSTEM
has detected fluid and
after 10 seconds
stopped all pumps
Error in the
EPROM ERROR
program memory
CALL SERVICE
English 120
Measures
If neither cleaning nor a lamp
change rectifies the error,
call customer service
Check sample feed
Call customer service
The device data must
be completely checked
(F1 key)
Call customer service
A correction is required
Rectify error, dry out system
and push F1 [MEAS]
(5 seconds) to start the system
again
When this error appears again
after switch-off and another
switch-on, call customer
service
Error in the
RAM ERROR
data storage
CALL SERVICE
CALIBRATION
Standard solution or
FACTOR
reagents missing
or are not
OK

Order numbers

Further accessories
Item no.
Designation
DOC023.XX.03107
User Manual
Available on request
FILTRAX with 2 m unheated hose
Available on request
FILTRAX with 10 m unheated hose
Available on request
FILTRAX with 20 m unheated hose
Available on request
FILTRAX with 30 m unheated hose
LZX414.00.00000
Tank edge hardware for module carrier
LZX676.00
Control unit bracket
LZX408
VIEWTAX program for data analysis
When this error appears again
after switch-off and another
switch-on, call customer
service
Check pumps, hoses and
solenoid valves
Check canister fill level

Quicklinks ausblenden:

Werbung

Inhaltsverzeichnis
loading