Herunterladen Inhalt Inhalt Diese Seite drucken

Hart-Command 44 : Write Primary Variable Units; Hart-Command 45 : Trim Primary Variable Current Dac Zero; Hart-Command 46 : Trim Primary Variable Current Dac Gain - ABB COPA-XE Betriebsanleitung

Vorschau ausblenden Andere Handbücher für COPA-XE:
Inhaltsverzeichnis

Werbung

Erstellt
von
Name:
Holger Seebode
Datum:
9.7.1998

4.6 HART-Command 44 : Write Primary Variable Units

Request
#0 Primary Variable Unit Code
Data Bytes
Response
#0 Primary Variable Unit Code
Data Bytes
Response
0
Codes
2
5

4.7 HART-Command 45 : Trim Primary Variable Current DAC Zero

Request
#0..#3 Externally Measured Primary Variable Current Level, IEEE 754, Units of mA
Data Bytes
Response
#0..#3 Actual Measured Primary Variable Current Level, IEEE 754, mA
Data Bytes
Response
0
Codes
3
4
5
9
11

4.8 HART-Command 46 : Trim Primary Variable Current DAC Gain

Request
#0..#3 Externally Measured Primary Variable Current Level, IEEE 754, Units of mA
Data Bytes
Response
#0..#3: Actual Measured Primary Variable Current Level, IEEE 754, mA
Data Bytes
Response
0
Codes
3
4
5
9
11
HART-Protokoll
Gerät: COPA/MAG-XE, 50XE4000
Software: HART-Software
Kennung: D699B180 X.10
Projektnummer:
No Command Specific Error
Invalid Selection
Incorrect Byte Count
No Command Specific Error
Passed Parameter to Large (> 5mA)
Passed Parameter to Small (< 3mA)
Incorrect Byte Count
Not in Proper Current Mode
In Multidrop Mode
No Command Specific Error
Passed Parameter to Large (> 22mA)
Passed Parameter to Small (< 18mA)
Incorrect Byte Count
Not in Proper Current Mode
In Multidrop Mode
Revision:
0
Name:
Holger Seebode
Datum:
9.7.1998
Revision
Revision
Revision
4-3