Rough Cleaning of Flow-Through-Chamber and Tubing

Flow-through-chamber and tubing might be roughly cleaned by pumping a suitable liquid through
the chamber instead of the medium.
9.4

Cleaning the Inside of the Flow-Through-Chamber

Before cleaning please disconnect first drive unit from mains and detach the flow through
chamber DK 30
The flow-through chamber can be disassembled without disconnecting the tubes. Attention: Rest
liquid escapes!!!
Please take care that neither O-rings nor any other parts experience any damage during
assembling respectively disassembling the flow-through chamber.
Attention: Before operation check the slip ring sealing. The slip ring should be easily turned by
hand. If it is stuck after a long period without use just loosen slip ring by wetting it with water.
Never operate the flow-through chamber if the slip ring is stuck!!
Note: Rotor and stator wrenches are listed in the price list under accessories. Always unscrew the
rotor before the stator.
9.5

Cleaning of the Analytical Mill Attachment AX 60

Before cleaning please disconnect first drive unit from mains and detach the mill attachment AX
60.
Please take care that neither the o-ring nor any other parts experience any damage during
assembling respectively disassembling the mill attachment.
Remove residues from the inside of the chamber. To avoid clogging, clean shaft, knife and the
inside of the chambers after each use. This is done by operating the attachment using a solvent
which dissolves substance residues and is not harmful to the gasket. This is usually sufficient to
clean the attachment inside.
Chemical sterilization may be also a method. General-purpose disinfectants such as formalin,
alcohol, etc. may be used. It is important to remove disinfectant residues with sterilized water.
Important Note:
Make sure that the bearings, O-rings and gaskets are resistant to solvents.
Sterilization with moist heat:
This denotes the use of a steam jet pressurized to 2 bars at 120°C.
9.6

Maintenance of drive units UNIDRIVE X1000/X1000D

Attention:
Do not open the instrument. Repairs are only to be carried out by trained service
technicians.
Instruction Manual UNIDRIVE X1000/X1000D 2011/08 V3.00