3.4

Installation procedure

The panel cutout required measures 92x45 mm (3.62x1.77 inches).
Fig. 1: Installation in the panel
1.
Screw the threaded rods (pos. 1) into the positions provided on the mounting frame (pos.
2). Four opposing screw positions (pos. 3/4) are available for this purpose.
2.
Push the device with the sealing ring through the panel cutout from the front.
3.
To secure the casing in the panel, hold the device in a horizontal position and push the
frame (pos. 2), with the threaded rods screwed in, over the casing until the frame locks
into position (1.).
4.
Then tighten the threaded rods to fix the device in place (2.).
5.
For the Ex option, install the distance piece (pos. 6) for the input terminals.
To disassemble the device, the mounting frame can be unlocked at the locking elements (pos.
5) and then removed.
3.5

Post-installation check

• Is the sealing ring undamaged?
• Is the mounting frame securely engaged on the housing of the device?
• Are the threaded rods tightened?
• Is the device positioned in the center of the panel cutout?
• Is the distance piece installed (Ex option)?