NM 3000
For care of the material, please do not fail to comply with the following important points:
Do not use any solvent-containing, caustic cleaning agents!
Do not clean devices or instruments with compressed air!
Only wipe control unit with cloth. Do not use spray cleaning, as device is not
sealed!

6.1 Motor or handpiece

See separate instructions for motor or handpiece!
Fault:
Device does not run:
Footpedal does not
function:
Micromotor 31 problems:
Handpiece 34 problems:
Handpiece 38 problems:
If a fault cannot be eliminated, please contact the supplier or the authorised service centre. Relevant
addresses can be found on the last page of the operating instructions.
When discarding the device, device components and accessories, please comply with the issued
statutory regulations.
With regard to the preservation of the environment old equipment may be returned to the distributor or
manufacturer.
6

6 Cleaning

7 Faults and fault detection

Cause:
Mains connector not plugged
Wrong operating voltage
Foot pedal not connected
Control unit not switched on
See separate operating instructions for micromotor 31
See separate operating instructions for handpiece 34
See separate operating instructions for handpiece 38

8 Discarding instructions

NOUVAG AG
Elimination:
Plug in mains plug
Check mains voltage
→See marking on device label
Connect foot pedal cable to back of
control unit
Switch "ON/OFF" main switch to "ON"
position
No. 31884 29/05