Herunterladen Inhalt Inhalt Diese Seite drucken

Removal Of Sources Of Malfunction; Disposal; Guarantee; Storage Conditions - Mettler Toledo InPro 42 Serie Betriebsanleitung

Ph-elektrode
Inhaltsverzeichnis

Werbung

Verfügbare Sprachen
  • DE

Verfügbare Sprachen

  • DEUTSCH, seite 10
InPro 42XX (i) Series pH electrode
3. If the electrode is stored in its housing the advices
described under Item 2 apply, although the
electrolyte has to be slightly modified for storage
purposes (2 parts of buffer solution pH 9.2 to
10  parts of electrolyte) in order to avoid any
corrosion of the housing parts.
4. If an electrode is inadvertently stored in a dry
condition for several days it must be rinsed
for several hours in a standard storage electrolyte.
5. The plug-in head should occasionally be checked
for possible traces of moisture. If necessary clean
it thoroughly with deionized water or alcohol and
then dry it off carefully.
7

Removal of sources of malfunction

With a soft toothbrush and water any residues can
easily be removed from the electrode. Mild washing
agents can also be used.
8

Disposal

It is the user's responsibility to dispose of the sensor
in a professional manner. The sensor can contain
electronic components that require disposal such that
no persons or the environment are endangered.
9

Guarantee

On manufacturing defects 12 months after delivery.
10

Storage conditions

We recommend that the electrodes are stored at room
temperature.
© 04 / 16 Mettler-Toledo GmbH
Printed in Switzerland
InPro 42XX (i) pH electrode
en
17
52 005 365

Werbung

Inhaltsverzeichnis
loading

Diese Anleitung auch für:

Inpro 42i serie

Inhaltsverzeichnis