Safety Classifications

CATEGORY
Protection Against Electric Shock
Protection Against Solid Object and
Liquid Ingress
Mode of Operation
* Type BF defibrillation proof applied parts are the underside of the chest pad and the top
of the back pad. See page 17 for symbol illustration.

Electromagnetic Compatibility Guidance

Electromagnetic Emissions
Table 1 Guidance and Manufacturer's Declaration - Electromagnetic Emissions
The TrueCPR coaching device, model 80596-000003, is intended for use in the
electromagnetic environment specified below. The customer or user should ensure that
the TrueCPR device is used in such an environment.
Emissions Test
RF emissions
CISPR 11
RF emissions
CISPR 11
Harmonic emissions
IEC 61000-3-2
Voltage fluctuations/
flicker emissions
IEC 61000-3-3
The TrueCPR device emits a continuous wave signal at 12 ± 0.1 kHz. The maximum
Effective Isotropic Radiated Power (EIRP) and Effective Radiated Power (ERP) levels are
-40 dBm, measured at a distance of 3 m (10 ft).
Essential Performance
The TrueCPR device maintains safe and effective performance (accurate depth) when
operated in the electromagnetic environment specified in Table 2 through Table 4 .
©2013 Physio-Control, Inc.
CLASSIFICATION
Internally powered ME equipment.
Type BF defibrillation proof applied parts.*
IP55
Continuous operation
Compliance
Electromagnetic Environment - Guidance
Group 1
The TrueCPR device uses RF energy only for its
internal function. Therefore, its RF emissions are
very low and are not likely to cause any interference
in nearby electronic equipment.
Class B
The TrueCPR device is suitable for use in all
establishments including domestic and those
directly connected to the public low-voltage power
Not applicable
supply network that supplies buildings used for
domestic purposes.
Not applicable
The TrueCPR device is battery-powered.