  • DEUTSCH, seite 1
– The measurement value flashes.
 Open the test gas cylinder valve to let gas flow over the sensor with a volume
flow of 0.5 l/min.
– The displayed, flashing measurement value changes to the value according
to the supplied test gas.
When the displayed measurement value is stable (after at least 120 seconds):
 Press the [OK] key to perform the calibration.
– The display containing the current gas concentration changes with the
display » OK «.
 Press the [OK] key or wait for approx. 5 seconds to end the calibration of this
measuring channel.
– The next measuring channel may appear for calibration.
– After the calibration of the last measuring channel, the instrument changes to
measuring mode.
 Close the test gas cylinder valve and remove the instrument from the
calibration cradle.
If a fault occurred during the span calibration:
– The fault message »
« appears and »
respective sensor instead of the measured value.
 In this case, repeat the calibration.
 If necessary, replace the sensor.
Notice for the adjustment of the ex-channel to nonane as a measuring gas:
When calibrating the Ex channel, propane can be used as a substitute for
the test gas.
When using propane to adjust the ex-channel to nonane, the display must
be set to twice the used test gas concentration.
Notice for use in underground mining:
When calibrating the Ex channel to methane as the measured gas,
the indication on the instrument must be set to a value 5 % (relative) higher
than the test gas concentration used.
Cleaning
The instrument does not need any special care.
 Dirt and deposits can be removed from the instrument by washing it with cold
water. A sponge can be used for wiping if necessary.
Abrasive cleaning tools (brushes etc.), cleaning agents and cleaning solvents
can destroy the dust and water filters.
 Carefully dry the instrument with a cloth.
« is displayed for the
NOTICE

Maintenance

Device maintenance must only be carried out by trained service personnel.
The instrument should be inspected and serviced once a year by suitably
qualified persons. Comparisons:
EN 60079-29-2 – Gas detectors - Selection, installation, use and
maintenance of detectors for flammable gases and oxygen
EN 45544-4 – Electrical apparatus used for the direct detection and direct
concentration measurement of toxic gases and vapours - Part 4: Guide for
selection, installation, use and maintenance
National regulations
Recommended calibration interval for measuring channels Ex, O
NO
and CO: 6 months. Calibration intervals of other gases: see Instructions for
2
Use of the applicable DrägerSensors.
See the Technical Manual for details of spare parts.
Storage
Dräger recommends storing the instrument in the charger module (order
no. 83 18 639).
Dräger recommends checking the charge of the power supply at least every
three weeks if the instrument is not stored in the charger module.
Disposal
This product must not be disposed of as municipal waste. This is
indicated by the adjacent icon.
You can return this product to Dräger free of charge. For information
please contact the national sales organisations and Dräger.
Batteries and rechargeable batteries must not be disposed of as
municipal waste. This is indicated by the adjacent icon. Collect and
dispose of batteries and rechargeable batteries at battery collection
centres, in accordance with applicable regulations.