Herunterladen Inhalt Inhalt Diese Seite drucken

PeakTech 4970 Bedienungsanleitung Seite 25

"3 in 1" infrarot-thermometer
Inhaltsverzeichnis

Werbung

Verfügbare Sprachen

Verfügbare Sprachen

5.5. Emissivity Adjustment Mode
The Emissivity Adjustable Mode useful for adjusting the emissivity.
Firstly turn the instrument on by pressing the "IRT" button, release
the "IRT" button to hold measuring data. Then press the "MODE"
button five times for the Emissivity Adjustable
Press the "▲" button or "IRT" button to adjust the Emissivity.
Most (90% of typical applications) organic materials and painted or
oxidized surfaces have an emissivity of 0.95 (pre-set in the unit).
Inaccurate readings will result from measuring shiny or polished
metal surfaces. To compensate, cover the surface to be measured
with masking tape or flat black paint.
Allow time for the tape to reach the same temperature as the
material underneath it. Measure the temperature of the tape or
painted surface.
5.6 Description of the distance to spot size
D =
Distance (spot size depends on the distance to the object) 1 :
1
S =
diameter of the spot center
-24-
Mode function.

Werbung

Inhaltsverzeichnis
loading

Inhaltsverzeichnis