 Ensure that there are no objects on the weighing pan. Press the TARE key.
The weight value of the required adjustment weight will be displayed.
 Carefully place the adjusting weight into the centre of the balancing plate and
press the TARE key.
After successful adjustment, an acoustic signal sounds, the scale will
automatically return to the menu. Take away adjustment weight.
Cancel the adjustment:
Press the BG/ NET key repeatedly, until
Press the BG/ NET key again, the display returns to the menu "02 EC".
KXS/KXG-TM-BA_IA-e-1621
„PASS"
is indicated.