Herunterladen Inhalt Inhalt Diese Seite drucken

Broken Glass; Faults; Acceptance Of The Centrifuges For Repair - Hettich UNIVERSAL 32 Bedienungsanleitung

Vorschau ausblenden Andere Handbücher für UNIVERSAL 32:
Inhaltsverzeichnis

Werbung

Verfügbare Sprachen

Verfügbare Sprachen

EN
23.4

Broken glass

In the event of broken glass the glass splinters and leaking centrifugal material are to be carefully removed from
the centrifugal chamber and from the containers or container drillings.
Replace the rubber inserts.
The rubber inserts of the container must be replaced after breakage of glass, because any remaining
glass splinters in the rubber cores could cause further breakage of glass.
If this concerns infectious material, a disinfection process is to be executed immediately.
24

Faults

If the fault cannot be rectified according to the faults table, Hettich customer services must be informed.
Please state the type of centrifuge and the factory serial number. Both values are visible on the centrifuge type plate.
MAINS RESET:
- Mains switch OFF for longer than 10 secs.
- Mains switch ON.
Message / fault
No display
TACHO - ERROR
IMBALANCE
CONTROL - ERROR
06 - 09
N > MAX
N < MIN
ROTORCODE
MAINS INTERRUPT
VERSIONS-ERROR
CONTROL-ERROR
21 - 27
SER I/O - ERROR
30 - 38
° C * - ERROR
50 - 56
LOCK - ERROR
FU / CCI - ERROR
60 - 83
N > ROTOR-MAX
25

Acceptance of the centrifuges for repair

If the centrifuge is returned to the manufacturer for repair, it must be decontaminated and cleaned to protect persons,
environment and material.
We reserve the right to accept contaminated centrifuges.
Costs incurred for cleaning and disinfection are to be charged to the customer.
We ask for your understanding in this matter.
You will find further information on our homepage: http://www.HettichLab.com
22/56
Cause
---
− No voltage.
− Overvoltage protection tripped out.
− Faulty speedometer.
01
02
− No rotor installed.
− Defective motor, frequency converter or
drive.
---
Imbalance about motor axis through weight
differential in rotor assembly.
04,
Error in lid locking or lid closure.
05
Rotation too fast
13
Rotation too slow
10
Incorrect rotor coding
---
Power failure, centrifuging not properly
completed
12
Mismatch between electronic components
Error / defective control unit
Error / defective interface
Error / defective cooling
57
Error / defective program locking
Error / defective motor control
---
Speed in the selected program greater than
the maximum speed of the rotor.
Remedy
− Check supply voltage.
− Mains switch ON.
− Open lid.
− Turn rotor manually.
− MAINS-RESET when power is
switched on, rotor should turn.
− Open lid.
− Correct imbalance.
− Open lid.
− MAINS-RESET
− Open lid.
− Push
button.
START
− Open lid.
− MAINS-RESET
− Check the set speed.
Reduce the set speed