VOLTCRAFT VC-440 E Bedienungsanleitung Seite 59

Inhaltsverzeichnis

Werbung

Verfügbare Sprachen

Verfügbare Sprachen

Direct voltage V/DC
Range
600.0 mV*
6.000 V
60.00 V
600.0 V
1000 V
*only available via the measuring function "mV"
Overload protection 1000 V; impedance: 10 MΩ (mV: ≥1000 MΩ)
When the measuring input is short-circuited in the mV area, a display of 5 counts is possible;
Accuracy range 5-100%
Alternating voltage V/AC
Range
600.0 mV*
6.000 V
60.00 V
600.0 V
750 V
600.0 V LoZ
*only available via the measuring function "mV"
Frequency range 40 - 400 Hz; overload protection 750 V; impedance: 10 MΩ
Specified measurement range: 5 - 100 % of the measured range
When the measuring input is short-circuited, a display of 10 counts is possible
After use of the LoZ function, a regeneration time of 1 minute is required
TrueRMS crest factor (CF) ≤3 CF to 600 V, ≤1.5 CF to 750 V
TrueRMS crest factor for non-sine-shaped signals plus tolerance addition:
CF >1.0 - 2.0 + 3%
CF >2.0 - 2.5 + 5%
CF >2.5 - 3.0 + 7%
Resolution
0.1 mV
0.001 V
0.01 V
0.1 V
1 V
Resolution
0.1 mV
0.001 V
0.01 V
0.1 V
1 V
0.1 V
Accuracy
±(0.8% + 3)
±(0.6% + 3)
±(0.8% + 3)
Accuracy
±(1.3% + 5)
±(1.0% + 5)
±(1.3% + 4)
±(2,6% + 4)
59