Herunterladen Inhalt Inhalt Diese Seite drucken

Measuring Temperature With A Resistance Thermometer - WIKA CPH6600 Betriebsanleitung

Druckkalibrator
Inhaltsverzeichnis

Werbung

Verfügbare Sprachen

Verfügbare Sprachen

6. Commissioning, operation

6.11 Measuring temperature with a resistance thermometer

For temperature measurement, select the function "RTD" on one of the displays.
The correct resistance thermometer must be used. There are four types of resistance
GB
thermometer supported: "P100-385", "P100-392", "P100-JIS" and "CUSTOM".
The WIKA Pt100 only needs to be connected to the CPH6600 and the measured
temperature will immediately be displayed.
If a custom resistance thermometer is being used, the entry of R0 and the coefficients is
handled through the serial interface.
34
The factory default type is "P100-385", so if the CPH6600 is being used
with a WIKA Pt100 you do not have to reset the resistance thermometer
type.
The display will indicate "OL" if the measured temperature is outside
the nominal measurement range of the resistance thermometer (below
-40 °C or above 150 °C).
Figure 3: CPH6600 with a resistance thermometer
WIKA operating instructions, model CPH6600

Werbung

Kapitel

Inhaltsverzeichnis
loading

Inhaltsverzeichnis