6. Commissioning, operation

6.8 Measuring pressure

For pressure calibration, connect the calibrator with an appropriate adapter. Select the
pressure setting for the display being used. The calibrator is equipped with an internal
GB
pressure sensor and optional external pressure modules are available. Choose a sensor
that is suitable for the pressure range and accuracy.

6.8.1 Media compatibility

The CPH6600 has a unique valve-cleaning port. This is to make the pump more
accessible for the user, and thus easier to maintain. Chapter 7.2 "Cleaning" describes
the cleaning of these valves. Even though servicing the pump is easy, care should be
taken that the calibrator is only used with clean, dry gases.
30
CAUTION!
Pressure sensors may be damaged and/or injuries may occur to
the personnel due to improper application of pressure. For a better
understanding with respect to overpressure and burst pressure, follow
the specifications laid down in these operating instructions (see chapter
3 "Specifications").
The calibrator display will indicate "OL" when an inappropriate pressure
is applied. As soon as "OL" is observed on any pressure display, the
pressure should be reduced immediately to prevent damage or possible
physical injury. "OL" is displayed if the pressure exceeds the nominal
range by 110 %.
Use the ZERO key to zero the pressure sensor once it is vented to
atmospheric pressure.
CAUTION!
The calibrator should only be used with clean, dry gases! To prevent
contamination by the test pieces, the use of a dirt trap is recommended
(see chapter 10 "Accessories")!
WIKA operating instructions, model CPH6600