REHA Programme
Prg.
PHASE 1
Total time 30 min
frequency 800 Hz
R1
impulse width 100µs
Total time 30 min
frequency 1000 Hz
R2
impulse width 100µs
Total time 30 min
frequency 1200 Hz
R3
impulse width 100µs
Total time 30 min
frequency 90 Hz
R4
impulse width 20µs
Total time 30 min (5 sec
30 Hz – 200 us + 5 sec
R5
50 Hz – 150 us + 5 sec
100 Hz – 120 us) x 120
cycles
Total time 30 min (6 sec
100Hz – 175 us + 6 sec
R6
2-100Hz modulated – 250
us + 6 sec 150Hz – 60-
200 us)
Total time 30 min (6 sec
100Hz – 175 us + 6 sec
R7
2-100Hz mod modulated
– 250 us + 6 sec 150Hz –
60-200 us modulated)
Total time 30 min
frequency 2 Hz
R8
impulse width 80 us
Burst impulses
Total time 4 min
frequency 6 Hz
R9
impulse width 250us
Total time 4 min
frequency 6 Hz
R10
impulse width 250us
IACER S.r.l.
PHASE 2
Total time 10 min
(10 sec 3Hz – 250us 80%
+ 5 sec 20Hz – 250us
80%) x 40 cycles
Total time 10 min
(10 sec 3Hz – 250us 80%
+ 5 sec 40Hz – 250us
80%) x 40 cycles
20 of 24
PHASE 3
Total time 10 min
(10 sec 3Hz – 250us
80% + 5 sec 30Hz –
250us 80%) x 40
cycles
Total time 10 min
(10 sec 3Hz – 250us
80% + 5 sec 50Hz –
250us 80%) x 40
cycles