Herunterladen Inhalt Inhalt Diese Seite drucken

Unitest BEHA CHB 5 Bedienungsanleitung Seite 38

Inhaltsverzeichnis

Werbung

Verfügbare Sprachen

Verfügbare Sprachen

Harmonics in mA, A (Absolute Value)
specified from 100 mA rms 50/60Hz (mA Range) or from 1A rms 50/60Hz (A Range),
max. Peak Voltage 850 mAss (mA Range) or 70 Ass (A Range)
Range
1...10
11...20
21...50
51...99
Harmonics in mA, A and V Measurement Range
(Total Harmonic Distortion THD %)
specified from 100 mA rms 50/60Hz (mA Range) or from 1 A rms 50/60Hz (A Range) or
from 80 V rms AC 50/60Hz (V Range), max. Peak voltage 850 mAss (mA Range) or 70 Ass
(A Range) or max. Peak Voltage 850 Vss
Range
0,0...10,0 %
10,0...40 %
40...100 %
100...999,9 %
PEAK Value in mA, A and V Measurement Range
specified from 10 mA rms (mA Range) or from 0,5 A rms (A Range) or from 20 V rms AC
(V Range), max. Peak voltage 850 mAss (mA Range) or 70 Ass (A Range) or max. Peak
Voltage 850 Vss
Range
50 Hz
60 Hz
Resolution
0,01 mA / 0,1 mA /
0,001 A / 0,01 A
Resolution
0,1 %
0,1 %
0,1 %
0,1 %
Sampling time
39 µs
33 µs
Accuracy (50/60Hz)
Display in mA: ± 1 % rdg. ± 2 mA
Display in A: ± 1 % rdg. ± 0,3 A
Display in mA: ± 7 % rdg. ± 2 mA
Display in A: ± 7 % rdg. ± 0,3 A
Display in mA: ± 15 % rdg. ± 3 mA
Display in A: ± 15 % rdg. ± 0,3 A
Display in mA: ± 35 % rdg. ± 3 mA
Display in A: ± 35 % rdg. ± 0,3 A
Accuracy
± 2 %
± 5 % rdg. ± 5 %
± 10 % rdg. ± 10 %
± 20 % rdg.
Accuracy
± 5 % rdg. ± 30 D
± 5 % rdg. ± 30 D
16

Quicklinks ausblenden:

Werbung

Kapitel

Inhaltsverzeichnis
loading

Inhaltsverzeichnis