4. Automatic zero point correction (Autozero) by changing the
display, digits selectable (0.5d, 1d, 2d, 4d)
Press
, the currently set value is displayed.
Press
to select the desired setting (0.5d, 1d, 2d, 4d) and
acknowledge by pressing
5. Zero setting range
Load range where the display after switching-on the balance is
set to zero
Press
, the currently set zero setting range is displayed.
Press
to select the desired setting (0 %, 2 %, 5 %, 10 , 20
%) and acknowledge by pressing
6. Zero setting range
Load range where the display is set to zero by pressing
Press
, the currently set auto-zero range is displayed.
Press
to select the desired setting (0 %, 2 %, 5 %, 10 %,
20 %) and acknowledge by pressing
44
.
.
.
CFS/CCS-BA-e-1215