Herunterladen Inhalt Inhalt Diese Seite drucken

Table Linearization Points - KERN CFS Betriebsanleitung

Vorschau ausblenden Andere Handbücher für CFS:
Inhaltsverzeichnis

Werbung

Verfügbare Sprachen

Verfügbare Sprachen

 Using
select the weighing unit [kg or lb]
which will be used for linearization. The appeared
[] displays the current weighing unit.
Acknowledge with
 Ensure that there are no objects on the weighing
pan. After successful standstill control press the
button.
 The weighing value of the first weight will be dis-
played.
 Place weight 1 carefully in the centre of the weighing
plate, wait for the stability display and confirm by
.
 The weighing value of the second weight will be dis-
played.
 Remove weight 1
 Place weight 2 carefully in the centre of the weighing
plate, wait for the stability display and confirm by
.
 After the adjustment the balance will carry out a self-
test. Remove adjusting weight during selftest, bal-
ance will return into weighing mode automatically.
In case of fault or a wrong weight an error message
(
earization procedure.
8.1.1

Table Linearization points

Max
Weight 1
6 kg
15 kg
„A 10 kg"
30 kg
22
.
) will appear, then repeat the lin-
Display
Weight
„A 2 kg"
2 kg
„A 5 kg"
5 kg
10 kg
(example)
(example)
Display
Weight
Weight 2
„A 6 kg"
6 kg
„A 15 kg"
15 kg
„A 30 kg"
30 kg