If necessary, at balance zero display using
select the weighing unit [kg or lb], which shall be
used for adjustment. The appeared [] displays the
current weighing unit.
Acknowledge with
Ensure that there are no objects on the weighing
plate. After successful stop check press the
ton.
The weight value of the required adjustment weight
will be displayed.
Acknowledge with
Place the required adjusting weight carefully in the
centre of the weighing plate and press
firm.
or
Enter value of the necessary adjustment weight
using numeric keyboard and acknowledge with
In order to achieve high-quality weighing results in
the sense of the measuring technology, it is recom-
mended to select the nominal value as high as poss-
ible.
We recommend 80 % max.
Place the required adjusting weight carefully in the
centre of the weighing plate and press
firm.
After the adjustment the balance will carry out a self-
test. Remove adjusting weight during self test, bal-
ance will return into weighing mode automatically.
In case of an adjustment error or incorrect adjusting
weight the display will show an error message
(
CFS/CCS-BA-e-1215
.
.
), repeat adjustment process.