7.2 Adjustment bulk weighing balance, not preconfigured by KERN

 Switch-on balance and during the self test press
.
 Use the number keys to enter password:
Either
Standard password „0000"
or
personal password, input see chapter 13
 Confirm input with
 Select bulk material balance using
The appeared [] displays the current balance.
Acknowledge with
 If necessary, at balance zero display using
select the weighing unit [kg or lb], which shall be
used for adjustment. The appeared [] displays the
current weighing unit.
Acknowledge with
 Ensure that there are no objects on the weighing
plate. After successful stop check press the
ton.
 Enter value of the necessary adjustment weight
using numeric keyboard and acknowledge with
In order to achieve high-quality weighing results in
the sense of the measuring technology, it is recom-
mended to select the nominal value as high as poss-
ible. We recommend 80 % max.
 Place the required adjusting weight carefully in the
centre of the weighing plate and press
firm.
 After the adjustment the balance will carry out a self-
test. Remove adjusting weight during self test,
balance will return into weighing mode automatically.
In case of an adjustment error or incorrect adjusting
weight the display will show an error message
(
20
Operation
.
.
.
.
), repeat adjustment process.
but-
.