3.2. Resistance
Range
Resolution
20
Ω
200
Ω
2 kΩ
20 kΩ
200 kΩ
100
2 MΩ
20 MΩ
200 MΩ
100
2000 MΩ
* When you short the test leads in the 200 Ω range, your meter
displays a small value (no more than 0.3 Ω). This value is due
to your meter's and test leads internal resistance. Make a note
of
this
value and subtract it
measurements for better accuracy.
3.3 Continuity and Diodes
Function
Diode
Continuity
Remarks:
Diode:
Continuity:
0,01 Ω
0,1 Ω
1
Ω
10
Ω
Ω
1
kΩ
10
kΩ
kΩ
1 MΩ
Range
. ))))
Leerlaufspannung: 2,8V
Audible continuity threshold: Less than
120 Ω
Accuracy
± 1,0% rdg. + 5dgt.
± 0,8% rdg. + 3dgt.
± 0,8% rdg. +1dgt.
± 1,0% rdg +
± 5.0% rdg. + 10dgt.
Reference only
from small resistance
Resolution
1mV
1 Ω
-22-
Overload
Protection
250V
2dgt.
Overload
Protection
250V
rms
rms