CAUTION
• Make sure that the tube is securely set on the
rollers when closing the Pump Cover. If the tube
is not correctly positioned on the Rollers and
the cover is closed, the tube could be cut or
sheared. (Fig.5)
• Do not mount the Irrigation Tube except the
person who specially educated.
• Do not mount the Irrigation Tube during Power is
ON.
After the tubes are correctly positioned, close the Pump Cover by
turning the Pump Cover Lever to the 'CLOSE' position. (180 degrees
counterclockwise) (Fig.6)
To remove, 'OPEN' the Pump Cover Lever then remove the tube.
4-6 Mounting the Coolant Solution Hanger Post
Mount the Coolant Solution Hanger Post onto the Holder on the
Control Unit. (Fig.7)
4-7 Insertion of the Irrigation Tube
1)
Close the Tube Clamp, between the Irrigation Tube Needle and the Irrigation Pump. (Fig.8)
2)
Insert the Irrigation Tube Needle into the Bottle Cap. (Fig.9)
3)
Open the Tube Cap to supply air into the bottle. (Fig.10)
4)
Open the Tube Clamp.
Tube Clamp
CAUTION
Do not operate the Irrigation Pump if the tube is bent or the Tube Clamp is in the closed position. This could cause
the tube to burst or slip out of the bottle.
Irrigation
Tube Needle
Fig.8
9
<CLOSE Position>
Bottle Cap
Tube Cap
Fig.9
Fig.5
Fig.6
Align
Fig.7
Fig.10