Herunterladen Inhalt Inhalt Diese Seite drucken

Finding Cable Breaks With 2 Transmitters; Finding Errors In Electrical Floor Heatings - VOLTCRAFT LSG-10 Bedienungsanleitung

Vorschau ausblenden Andere Handbücher für LSG-10:
Inhaltsverzeichnis

Werbung

Verfügbare Sprachen

Verfügbare Sprachen

Finding Cable Breaks with 2 Transmitters

Switch off the fuse and power down the electrical
circuit.
This method makes it possible to locate the error
from two sides. A different signal code is set for each
transmitter. The receiver can be used to easily
determine the respective signal direction. The trans-
mitter number is indicated accordingly.
Advantage: The two search signals will not interfere
with each other.
When you have found the location of the fault, the
receiver will no longer show any signal code, becau-
se both signals have the same strength there.
The cable break must be high-impe-
dance (>100 kOhm).

Finding Errors in Electrical Floor Heatings

Ensure that there is no insulating foil / mat with an
earth connection above the heating wires. If requi-
red, disconnect this earth connection before mea-
suring, or the error site will not be easy to locate.
The signal should be fed from both sides. It is more
comfortable to use a second transmitter unit with a
different signal code.
39

Quicklinks ausblenden:

Werbung

Inhaltsverzeichnis
loading

Diese Anleitung auch für:

12 19 44

Inhaltsverzeichnis