Calibration
The thermometer is calibrated at the time of manufacture.
If at any time you question the accuracy of temperature measurements,
please contact an authorized Kaz Service Center.

Product specifications

Model:
Displayed temperature range:
Operating ambient
temperature range:
Storage temperature:
Display resolution:
Accuracy for displayed
temperature range:
Clinical repeatability:
Operating and storage
relative humidity:
Battery life:
Service life:
If device is not used within specified temperature and humidity ranges
the technical accuracy cannot be ensured.
Type BF equipment
Storage temperature
Operating temperature
12
IRT 3020
34 °C – 42.2 °C
(93.2 °F – 108 °F)
10 °C – 40 °C
(50 °F – 104 °F)
-25°C – 55°C
0.1 °C or °F
± 0.2 °C (35.5 – 42 °C)
(95.9 – 107.6 °F)
± 0.3 °C (outside this
temperature range)
± 0.25 °C (± 0.45 °F)
≤ 95% non-condensing
2 years /
1000 measurements
5 years
See instruction of use
Keep it Dry
Subject to change without notice.
Standard Reference
Title
EN 12470-5
Clinical Thermometers – Part 5:
Performance of Infrared Ear
Thermometers (with maximum device).
EN 60601-1
Medical electrical equipment – Part 1:
General requirements for basic safety
and essential performance.
EN 60601-1-2
Medical electrical equipment – Part1-2:
General requirements for basic safety
and essential performance
– Collateral standard:
Electromagnetic compatibility –
Requirements and tests.
This product conforms to the provisions of the EC directives 93/42/EEC
(Medical Device Directive) and 2011/65/EU (RoHS).
MEDICAL ELECTRICAL EQUIPMENT needs special precautions
regarding EMC. For detailed description of EMC requirements please
contact an authorized local Service Centre.
(See Guarantee Card).
Portable and mobile RF communications equipment can affect
MEDICAL ELECTRICAL EQUIPMENT.
Please do not dispose of the product in the household
waste at the end of its useful life. Disposal can take place at
your local retailer or at appropriate collection points provided
in your country.